Worcester Polytechnic Institute Electronic Projects Collection

Title page for E-project-042507-193055


Project TypeMQP
Submission date2007-04-25
Authors
  • Michael Coutts, BE
  • Robert D Lavado, BE
  • Justin Parker, ME
  • URNE-project-042507-193055
    SponsorJohn Wixted, M.D., UMass Medical School
    TitleProtocol for Evaluating the Dynamic Compression of Knee and Ankle Arthrodesis Nails
    Advisor
  • Billiar, Kristen, BE
  • Availability unrestricted

    Abstract

    It is important for an arthrodesis nail to maintain compression for successful bony fusion. To test new designs of arthrodesis nails a device and protocol were developed to accurately measure compression within a joint over time. The protocol utilized a uniaxial testing machine to cyclically load the samples 10,000 times at 1 Hz from 0-200 lbs. Using Sawbones as a testing model, a 10% decrease in compression was experienced with a standard locked nail. Using this method, new and old designs of arthrodesis nails can be compared to determine which best maintains compression.