Robotically Controlled Magnetic Laser Aiming for Endoscopic Procedures
Abstract: We present a novel optical fiber steering system for laser surgery based on magnetic actuation. Fiber steering is achieved via two permanent magnets: a small magnet mounted at the fiber’s distal tip and a larger, robotically manipulated magnet that generates controlled magnetic torque and force. In clinical use, the larger magnet would remain outside the patient’s body, positioned close enough to produce the necessary magnetic fields, and the optical fiber would extend from an endoscope’s instrument channel. We first test magnetic laser aiming control and report tracking errors of 1.0 mm (open-loop) and 0.2 mm (closed-loop) over a 1 cm x 1 cm working area. We then pair the fiber with a clinical endoscope and demonstrate the system in a laryngeal phantom.
Advisor: Professor Loris Fichera
Committee: Professor Gio Pittiglio, Professor Connor McCann
Zoom Link: https://wpi.zoom.us/j/97537145033